SPECS

The SPECS Technologies Picoammeter

New femto High Frequency Upgrade!

LowNoise High Frequency Upgrade for Current Amplification

General Specifications:

Current Range: 2 nA - 2 mA

Resolution:

100 fA
Input Voltage Burden: < 200 µV on all current ranges
Display: Digital panel meter with 4 ½-digit resolution
Display Reading Rate: 3/s
Ranging: Manual or remote controlled (optional)
Overrange Indication: Display flashes "0000"
Current Source Bias: Push-button selectable: -90V, 0V, +90 V
Analog Output: Scaled voltage output, 2V full scale on all ranges
Zero Adjustment: Through two potentiometers on rear panel
Common Mode Voltage: Maximum 90 V
Input Connector: BNC on front panel
Input Protection: Overvoltage and arc protected current input
Isolation: Meter common to chassis: typically > 100 GOhm
Dimensions: 130 mm high x 110 mm wide x 170 mm deep
Weight: 1.5 kg
Power: 100-240V, 50-60Hz, 10VA


Remote Interface Specification (optional):

Remote Connector: 9-pin Dsub connector on rear panel
Input Signals: 3 TTL-signals for range selection
  1 TTL-signal for selecting local or remote control
Output Signals: 3 TTL-signals for reading the selected range
  1 analog signal for current monitoring (0-2V or 0-10V)

Instrument Accuracy:

Range
Max
Display
Accuracy (1 day)
Temp Coefficient
Analog
.
Range
Resolution
+-(%Reading + %Range) @ 25°C
+-(%Reading + %Range)/°C
Rise Time
pA
1999.9 pA
0.1pA
0.5 % + 0.05%

0.02% + 0.002%
100 ms
nA
199.99 nA
10 pA
0.1 % + 0.05%
0.002% + 0.002%
100 ms
uA
19.999 uA
1 nA
0.1 % + 0.05%
0.002% + 0.002%
10 ms
mA
1.9999 mA
100 nA
0.1 % + 0.05%
0.002% + 0.002%
1 ms
 

Digital Panel Meter Accuracy: +/- 1 digit , temperature drift: 0.4 counts/°C

Analog Output Accuracy: 2V full scale on all ranges, accuracy 1% of reading + 2mV offset



SPECS Technologies Corporation
3318 Plantation Dr.
Sarasota, Florida 34231
Phone: (941) 362-4877
Fax: (941) 364-9706
support@specs.com

News | Home | Products | Site Map | Contact Info | E-Mail Us