The EA 120 MCD is designed for use in an ESCA system to perform XPS, SSXPS (Small Spot XPS), UPS, AES, SAM and ISS measurements. These Electron/Ion Spectroscopy techniques, in particular XPS and Small Spot XPS, offer the possibility to obtain quantitative elemental and chemical information from surface layers, providing a powerful analysis tool for a wide range of materials.
| Method | Excitation | Line | Signal (cps) | Note |
|---|---|---|---|---|
| XPS
dE=const. |
Mg
300 W 15 keV |
Ag 3d 5/2 | 2.500.000 | FWHM
1.0 eV |
| Small
Spot XPS dE=const. |
Mg
300 W 15 keV |
Ag 3d 5/2 | 20.000 | FWHM
1.40 eV and 1000 µm accep. area |
| MXPS
dE=const. |
monochromatized
Al Kalpha 300 W,15 keV |
Ag 3d 5/2 | 40.000 | FWHM
0.65 eV |
| ISS
dE/E=3 |
He+,
2 keV, 1 µA biased +90V |
Ag
elastic peak |
1.000.000 | differentially
pumped fine focus Ion Source |
| UPS > dE/E=3 | He
I He II |
Au 5d band | 1.000.000 250.000 |
Gas Discharge UV Lamp |
The EA 120 MCD is a Multi-Channel Detector (MCD) version using a five-channeltron array to record energy spectra for negative charged particles (electrons) and positive charged particles (ions). 6 different entrance slits (selectable, indexed) allow the choice of various spot sizes and solid angles of acceptance. The analyzer uitilizez a Fringe Field Correction for Ultimate Energy Resolution.
Specifications EA 120 MCD:
| Hemispherical Electrostatic Capacitor | |
| Mu-metal shielding is built into the UHV housing for the analyzer and input lens system | |
|
Input lens system includes a first lens stage (non retarding = constant
analysis area for all kinetic energies) with variable angular aperture
and image aperture and a second lens stage (retarding) |
|
| Mean radius | 120 mm |
| Angle of spheres | 180° |
| Spatial resolution | 1000 µm for Small Spot XPS |
| Acceptance area | variable 1000 µm to 5 x 10mm |
| Acceptance angle | variable 4 to 12 degree |
| Flange to sample distance | 196 mm (others optional) |
| Working distance | 40 mm |
| Bakeable up to | 250 °C |
| Mounting flange | NW 100 CF |
The analyzer is controlled by the SPECTRA Data Acquisition and Processing System:
Please
download your EA 120 MCD Energy Analyzer Brochure ![]()
The EA 120 MCD is designed for use in OmniVac SPECS ESCA Systems:
Please
download your XPS Systems Brochure ![]()